Objectives: To study the clinical value of Dual-Energy Virtual Non-Contrast CT scan for subarachnoid hemorrhage. Methods: 30 patients of subarachnoid hemorrhage were selected, they received Dual-Source Conventional Non-Contrast CT scans and head angiography with Dual-Energy CT. Then Liver Virtual Non-Contrast program was used to reconstruct the images. After that the Conventional Non-Contrast images were taken as gold standard to evaluate the clinical value of Virtual Non-Contrast technology in the following aspects: Signal-Noise Ratio, Radiation Dose and lesion diagnosis. Results: The Signal-Noise Ratio of Virtual Non-Contrast CT scan was (4.21 ± 0.5), the Signal-Noise Ratio of Conventional Non-Contrast CT scan was (10.39 ± 0.71), the latter was significantly higher than the former. The Dose Length Product of Virtual Non-contrast CT scan was (121.30±3.8), it was close to 1/6 of the Dose Length Product (697.6±4.2)of the Conventional Non-Contrast CT scan. As for the lesion area, the difference of the CT values between the Virtual Non-Contrast and the Conventional Non-Contrast CT scan was only 4~10Hu. In addition, the Virtual Non-Contrast images’ quality was graded by diagnostic teachers: there were 4 cases of 3 points, 23 cases of 2 points, 3 case of 1 point. Conclusions: Compared with the Conventional Non-Contrast images, there is greater noise in the Virtual Non-Contrast images, but the latter can meet the clinical diagnostic requirements and greatly reduce the radiation dose. Advances in knowledge: Virtual Non-Contrast technology was used for the diagnoses of subarachnoid hemorrhage, and the radiation dose was greatly reduced.
Published in | Advances in Bioscience and Bioengineering (Volume 1, Issue 3) |
DOI | 10.11648/j.abb.20130103.11 |
Page(s) | 49-52 |
Creative Commons |
This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited. |
Copyright |
Copyright © The Author(s), 2014. Published by Science Publishing Group |
Dual-Source CT, Virtual Non-Contrast CT Scan, Signal-Noise Ratio, Radiation Dose
[1] | Jiang Jie, Han Dan. The clinical application and progress of virtual non-contrast technology of dual-energy CT. Chinese Journal of Radiology. 2012,46(10):958~960 |
[2] | Liang Yi, Dual-energy CT virtual non-contrasted images in patients suspected of having hepatic masses. Master’s Degree thesis, Southern Medical University, 2010. Page: 25-37. |
[3] | Guo Xing, Ding Wei. Application of virtual non-enhanced images in evaluating subarachnoid hemorrhage by dual-energy Computed Tomography angiography. Acta Academiae Medicinae Sinicae. 2010,Vol.32 No.6:695~698 |
[4] | Huang Wei, Xu Yiming. Dual-source virtual non-contrast CT of the head: a preliminary study. Chinese Journal of Radiology. 2011,Vol.45, No.3:229~233 |
[5] | Zhan Libo, Qin Dongjing. The basic principle and characteristics of dual-source CT. Radiologic Practice. 2008;06: 55~59 |
[6] | Lu Heqin, Zhu Guoying. To study the influencing factors of medical X-ray CT radiation dose. Shanghai Medical Imaging, 2008;02: 79~82 |
[7] | Zhang longjia, Lu Guangming. The research progress of dual-energy imaging of dual-source CT. Journal of Clinical Radiology, 2009;11:22~26 |
APA Style
Bo Xiao. (2014). To Explore the Clinical Value of Dual-Energy Virtual Non-Contrast CT Scan for Subarachnoid Hemorrhage. Advances in Bioscience and Bioengineering, 1(3), 49-52. https://doi.org/10.11648/j.abb.20130103.11
ACS Style
Bo Xiao. To Explore the Clinical Value of Dual-Energy Virtual Non-Contrast CT Scan for Subarachnoid Hemorrhage. Adv. BioSci. Bioeng. 2014, 1(3), 49-52. doi: 10.11648/j.abb.20130103.11
AMA Style
Bo Xiao. To Explore the Clinical Value of Dual-Energy Virtual Non-Contrast CT Scan for Subarachnoid Hemorrhage. Adv BioSci Bioeng. 2014;1(3):49-52. doi: 10.11648/j.abb.20130103.11
@article{10.11648/j.abb.20130103.11, author = {Bo Xiao}, title = {To Explore the Clinical Value of Dual-Energy Virtual Non-Contrast CT Scan for Subarachnoid Hemorrhage}, journal = {Advances in Bioscience and Bioengineering}, volume = {1}, number = {3}, pages = {49-52}, doi = {10.11648/j.abb.20130103.11}, url = {https://doi.org/10.11648/j.abb.20130103.11}, eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.abb.20130103.11}, abstract = {Objectives: To study the clinical value of Dual-Energy Virtual Non-Contrast CT scan for subarachnoid hemorrhage. Methods: 30 patients of subarachnoid hemorrhage were selected, they received Dual-Source Conventional Non-Contrast CT scans and head angiography with Dual-Energy CT. Then Liver Virtual Non-Contrast program was used to reconstruct the images. After that the Conventional Non-Contrast images were taken as gold standard to evaluate the clinical value of Virtual Non-Contrast technology in the following aspects: Signal-Noise Ratio, Radiation Dose and lesion diagnosis. Results: The Signal-Noise Ratio of Virtual Non-Contrast CT scan was (4.21 ± 0.5), the Signal-Noise Ratio of Conventional Non-Contrast CT scan was (10.39 ± 0.71), the latter was significantly higher than the former. The Dose Length Product of Virtual Non-contrast CT scan was (121.30±3.8), it was close to 1/6 of the Dose Length Product (697.6±4.2)of the Conventional Non-Contrast CT scan. As for the lesion area, the difference of the CT values between the Virtual Non-Contrast and the Conventional Non-Contrast CT scan was only 4~10Hu. In addition, the Virtual Non-Contrast images’ quality was graded by diagnostic teachers: there were 4 cases of 3 points, 23 cases of 2 points, 3 case of 1 point. Conclusions: Compared with the Conventional Non-Contrast images, there is greater noise in the Virtual Non-Contrast images, but the latter can meet the clinical diagnostic requirements and greatly reduce the radiation dose. Advances in knowledge: Virtual Non-Contrast technology was used for the diagnoses of subarachnoid hemorrhage, and the radiation dose was greatly reduced.}, year = {2014} }
TY - JOUR T1 - To Explore the Clinical Value of Dual-Energy Virtual Non-Contrast CT Scan for Subarachnoid Hemorrhage AU - Bo Xiao Y1 - 2014/01/30 PY - 2014 N1 - https://doi.org/10.11648/j.abb.20130103.11 DO - 10.11648/j.abb.20130103.11 T2 - Advances in Bioscience and Bioengineering JF - Advances in Bioscience and Bioengineering JO - Advances in Bioscience and Bioengineering SP - 49 EP - 52 PB - Science Publishing Group SN - 2330-4162 UR - https://doi.org/10.11648/j.abb.20130103.11 AB - Objectives: To study the clinical value of Dual-Energy Virtual Non-Contrast CT scan for subarachnoid hemorrhage. Methods: 30 patients of subarachnoid hemorrhage were selected, they received Dual-Source Conventional Non-Contrast CT scans and head angiography with Dual-Energy CT. Then Liver Virtual Non-Contrast program was used to reconstruct the images. After that the Conventional Non-Contrast images were taken as gold standard to evaluate the clinical value of Virtual Non-Contrast technology in the following aspects: Signal-Noise Ratio, Radiation Dose and lesion diagnosis. Results: The Signal-Noise Ratio of Virtual Non-Contrast CT scan was (4.21 ± 0.5), the Signal-Noise Ratio of Conventional Non-Contrast CT scan was (10.39 ± 0.71), the latter was significantly higher than the former. The Dose Length Product of Virtual Non-contrast CT scan was (121.30±3.8), it was close to 1/6 of the Dose Length Product (697.6±4.2)of the Conventional Non-Contrast CT scan. As for the lesion area, the difference of the CT values between the Virtual Non-Contrast and the Conventional Non-Contrast CT scan was only 4~10Hu. In addition, the Virtual Non-Contrast images’ quality was graded by diagnostic teachers: there were 4 cases of 3 points, 23 cases of 2 points, 3 case of 1 point. Conclusions: Compared with the Conventional Non-Contrast images, there is greater noise in the Virtual Non-Contrast images, but the latter can meet the clinical diagnostic requirements and greatly reduce the radiation dose. Advances in knowledge: Virtual Non-Contrast technology was used for the diagnoses of subarachnoid hemorrhage, and the radiation dose was greatly reduced. VL - 1 IS - 3 ER -